Incoming Material Inspection
Check sockets, Bourdon tubes, movements, dials, cases, seals and damping oil for material, dimensions, threads, appearance and cleanliness before assembly.
INDUSTRIAL PRESSURE GAUGES FROM DONGYA INSTRUMENT
Hangzhou Dongya Instrument supplies standard pressure gauges for process monitoring, pumps, compressors and hydraulic equipment. Select Y-60, Y-60ZT or Y-100ZT configurations with matched ranges, connections and mounting options; OEM dial and specification adjustments are available.
RANGE AND ACCURACY
Use the dial size and accuracy class to balance readability, installation space and measurement tolerance before selecting the pressure range.
| Starting Option | Concrete Specification | Best Fit |
|---|---|---|
| Y-60 / Y-60ZT | 60 mm dial, Accuracy Class 2.5, M14×1.5 | Compact panels and routine local indication |
| Y-100ZT | 100 mm dial, Accuracy Class 1.6, M20×1.5 | Stand-off viewing and tighter stated accuracy |
| Range | 0–0.6, 0–1.6 and 0–2.5 MPa certified configurations | Select the smallest range covering normal and foreseeable peak pressure |
Recommendation: Start with Y-60 or Y-60ZT for compact general measurement points and Y-100ZT for larger, more visible or Class 1.6 requirements; include normal, maximum and transient pressure in the quote.
CONNECTION AND MOUNTING
Thread standard, outside diameter, pitch and sealing method must match the equipment port; the dial and mounting direction then determine how easily operators can read and service the gauge.
| Installation Condition | Recommended Option | Why It Matters |
|---|---|---|
| 60 mm compact point | M14×1.5 with Y-60 or Y-60ZT | Keeps the compact gauge aligned with the listed Y-series connection |
| 100 mm or larger gauge point | M20×1.5 with Y-100ZT | Matches the standard connection used for Y-100 and larger Y-series gauges |
| Panel or board viewing | Axial or bezel mounting | Faces the dial toward the operator; radial mounting suits direct pipe takeoffs |
Recommendation: Send the equipment port thread, dial diameter, mounting direction and sealing method with the inquiry. Use the standard Y-series pairing first and request a custom connection only when the installation drawing requires it.
DELIVERY PATH
The fastest path is a listed dial, range and connection that fits the equipment. Custom production adds control for special threads, scales, filling, markings or documents.
Recommendation: Put the complete technical and document specification in the first inquiry so Dongya can return one matched configuration, quotation basis and delivery schedule before production.
QUALITY CONTROL, STANDARDS & CERTIFICATIONS
Each pressure gauge follows incoming inspection, assembly checks, five-point calibration and final leak testing. Current management-system and product type-approval evidence is shown below; third-party reports are arranged per order.
Check sockets, Bourdon tubes, movements, dials, cases, seals and damping oil for material, dimensions, threads, appearance and cleanliness before assembly.
Verify gear engagement, pointer zero setting, 0.3–0.8 mm pointer-to-dial clearance, sealing-ring position and fill quality during assembly.
Run five-point ascending and descending checks at 0%, 20%, 50%, 75% and 100% FS, then record indication error, hysteresis, tapping displacement, smooth movement and 1.25×FS leak-test results.
ISO 9001:2015 covers the pressure gauge manufacturing quality management system. Product type-approval certificates apply only to their listed models and ranges.
PRODUCT CATALOG
Compare the three specified standard pressure gauge models by dial size and connection. M14×1.5 is used for the listed 60 mm gauges; M20×1.5 is used for Y-100ZT. Request your working range and mounting orientation for a matched configuration.
Try another tag or contact us for a custom specification.
Request a recommendation ↗APPLICATION SCENARIOS
Y-series dial gauges provide local pressure indication on water, air, steam and non-corrosive liquid or gas lines when no explosion hazard or crystallizing medium is present.
Use a readable dial and a range that covers normal pressure, startup variation and foreseeable peaks when monitoring pump discharge, compressor receivers or hydraulic units.
Choose the 60 mm models for compact points and the 100 mm Y-100ZT when the dial must be read from farther away or the installation provides more panel space.
FAQ
Technical and commercial answers for selecting the listed pressure gauge configurations.
The supplied Y-series range includes 60 mm and 100 mm dial models. Certified configurations include 0–0.6, 0–1.6 and 0–2.5 MPa, with additional ranges selected by model. Choose a range that keeps normal pressure in the middle portion of the scale and leaves room for startup and pulsation.
Provide the medium, pressure basis, normal and maximum pressure, dial diameter, radial or axial mount, thread, case and wetted materials, temperature, filling and quantity. The listed 60 mm models use M14×1.5 and Y-100ZT uses M20×1.5 as the standard connection; custom threads and scales follow the order specification.
Standard in-stock gauges normally dispatch in 1–2 working days after order confirmation. Standard custom gauges normally require 3–5 working days for range, thread, fill, dial or mounting changes; quantity, special fittings and testing can extend scheduling. Products use protective film, shock-absorbing foam and cartons, with model and range marks plus a conformity certificate.
Each pressure gauge receives factory full inspection covering appearance, zero, rising and falling indication, hysteresis, pointer movement and a 1.25×FS leak test. ISO 9001:2015 certificate 41624QT7363R0 is listed through 2027-11-25, while product type-approval certificates apply only to their listed models and ranges. Third-party reports are arranged by advance request.
REQUEST A TECHNICAL QUOTE
Send the operating medium, normal and maximum pressure, dial size, mounting direction, thread, material, quantity and document needs. Dongya will return a matched configuration and delivery basis.